This guide will show you how to access and replace the battery of the ventilator.

Preparation to service:

  1. Make sure anaesthetic machine is OFF
  2. Disconnect the machine from power supply
  3. Gain access to ventilator housing from guide below

For more information on replacing the battery, please refer to page 29 in the service manual.

    • Before disassembly, make sure the front wheel breaks are on.

    • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

  1. Ulco Integrus PSV Battery Replacement: paso 4, imagen 1 de 1
    • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

  2. Ulco Integrus PSV Battery Replacement: paso 5, imagen 1 de 1
    • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

  3. Ulco Integrus PSV Battery Replacement, Removing the Rear Access Panels in the Ulco Integrus PSV: paso 6, imagen 1 de 1
    • Remove the rear plastic cover by turning the single hex key retaining clip to free the cover.

  4. Ulco Integrus PSV Battery Replacement: paso 7, imagen 1 de 1
    • Remove the four screws mounted to the cover plate using a 4mm Allen key.

  5. Ulco Integrus PSV Battery Replacement, Battery: paso 8, imagen 1 de 1
    • Remove the main cover plate from the ventilator enclosure by removing all screws.

    • Disconnect the battery.

  6. Ulco Integrus PSV Battery Replacement: paso 10, imagen 1 de 1
    • Remove the four pan head screws (SM0308P) holding down the battery plate (EV9109) and battery (EV9108) and remove the battery plate and battery.

  7. Ulco Integrus PSV Battery Replacement: paso 11, imagen 1 de 1
    • Install the new battery in the same location and screw down the battery plate onto the battery plate spacers so that the battery is firmly held in place.

    • Reconnect the battery.

Conclusión

To reassemble your device, follow these instructions in reverse order.

Preston Myers

Miembro Desde 04/09/20

493 Reputación

4 Guías creadas

Equipo

Cal Poly, Team S7-G13, Paton Spring 2020 Miembro de Cal Poly, Team S7-G13, Paton Spring 2020

CPSU-PATON-S20S7G13

3 Miembros

8 Guías creadas

0 comentarios

Agregar Comentario

Ver Estadísticas:

Ultimas 24 horas: 0

Ultimos 7 días: 0

Ultimos 30 días: 0

Todo El Tiempo: 51