Fisher Paykel Neopuff Valve System Replacement

Fisher Paykel Neopuff Valve System Replacement

Constantine Yap

Constantine Yap y 2 otros contribuidores

Última actualización: July 3, 2020

15 minuto
Moderado
Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.

Introducción

Ir al paso 1

NOTE: The valves are an integral part of the valve, panel and manifold assembly and are not able to be serviced. Please specify the model number from the parts list (section 4.1.2) when ordering a replacement valve assembly.

Please reference the parts list below for the disassembly diagram.

Please refer to page 7 of the device’s Technical Manual for source information.


Qué necesitas

  1. Fisher Paykel Neopuff Valve System Replacement, Valve System: paso 1, imagen 1 de 1
    • Remove the back cover, fixed by four screws.

    • Disconnect the tube from the manometer.

    • Remove the manometer by unscrewing the two retaining nuts.

    • Fit the existing manometer into the new valve assembly panel.

    • Tighten the retaining nuts.

    • Reconnect the manometer tube.

    • Refit the front panel to the back cover with the four screws from step 1.

  2. Fisher Paykel Neopuff Valve System Replacement: paso 8, imagen 1 de 1
    • Carry out the valve system performance test described above. It is recommended to record the lot number of the new valve assembly onto the Set-up and Maintenance Checklist.

Conclusión

To reassemble your device, follow these instructions in reverse order.

Constantine Yap

Miembro Desde 05/04/20

364 Reputación

2 Guías creadas

Equipo

Cal Poly, Team S1-G6, Zimmerman Spring 2020 Miembro de Cal Poly, Team S1-G6, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G6

Miembros de 4

27 Guías creadas

0 comentarios

Agregar Comentario

Ver Estadísticas:

Ultimas 24 horas: 2

Ultimos 7 días: 2

Ultimos 30 días: 15

Todo El Tiempo: 233