Cleaning the Ulco Integrus PSV

Cleaning the Ulco Integrus PSV

Jacob Henriksen

Jacob Henriksen y 2 otros contribuidores

Última actualización: July 3, 2020

15 minuto
Moderado
Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.

Introducción

Ir al paso 1

Step-by-step guide to take apart and clean absorber and bellows assembly for an ULCO Integrus PSV ventilator.

To begin, make sure the ventilator is turned off and unplugged.

These should be cleaned each day that the ventilator is in use (usually at the end of the day).

NOTE: Screw types and lengths are not specified in manuals

  1. Cleaning the Ulco Integrus PSV, Cleaning the Ulco Integrus PSV: paso 1, imagen 1 de 2 Cleaning the Ulco Integrus PSV, Cleaning the Ulco Integrus PSV: paso 1, imagen 2 de 2
    • Absorber

    • Loosen the knurled screws and remove the clear lid.

  2. Cleaning the Ulco Integrus PSV: paso 2, imagen 1 de 1
    • Remove the silicone shutters by gently lifting from the base, NOT from the flaps.

  3. Cleaning the Ulco Integrus PSV: paso 3, imagen 1 de 1
    • To release and remove the canister(s), undo the handle latch by turning sideways.

  4. Cleaning the Ulco Integrus PSV: paso 4, imagen 1 de 1
    • Push the release handle down to vertical until canisters are free.

  5. Cleaning the Ulco Integrus PSV: paso 5, imagen 1 de 1
    • Push canister(s) down then out of the cradle. Then dispose of the used soda lime (see suppliers recommended procedure for disposal of soda lime).

    • Wash all the absorber components. Do NOT use caustic cleaning fluids.

    • The Canister(s) and silicone shutters can be autoclaved.

    • The Absorber head can be put through a washer at 80 deg.C

    • Dry thoroughly before assembly, low pressure warm air should be passed through the head by attaching a hose to the expiratory port of the absorber.

  6. Cleaning the Ulco Integrus PSV: paso 6, imagen 1 de 2 Cleaning the Ulco Integrus PSV: paso 6, imagen 2 de 2
    • Bellows Assembly

    • Loosen the four knurled screws (1) and remove the bellows canister.

  7. Cleaning the Ulco Integrus PSV: paso 7, imagen 1 de 1
    • Remove the bellows assembly by rotating counter-clockwise. The bellows canister may be washed or autoclaved.

  8. Cleaning the Ulco Integrus PSV: paso 8, imagen 1 de 1
    • The base disk (mushroom) should be removed from the bellows assembly before washing. Pull the latex-free rubber bag from the delrin base disk. The bellows may be washed or autoclaved. The base component (mushroom) should not be autoclaved but can be put through a washer at 80°C.

    • Dry all components thoroughly before re-assembly.

    • Low pressure warm air should be passed through the absorber head by attaching a hose to the scavenge port

Conclusión

To reassemble your device, follow these instructions in reverse order.

Una persona más ha completado esta guía.

Equipo

Cal Poly, Team S7-G13, Paton Spring 2020 Miembro de Cal Poly, Team S7-G13, Paton Spring 2020

CPSU-PATON-S20S7G13

Miembros de 3

8 Guías creadas

0 comentarios

Agregar Comentario

Ver Estadísticas:

Ultimas 24 horas: 0

Ultimos 7 días: 2

Ultimos 30 días: 3

Todo El Tiempo: 98