SIREMOBIL Iso-C

SP Installation and Setting Instructions Connecting a laser camera © Siemens AG 1998 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
Installation and Setting Instructions
Connecting a laser camera
© Siemens AG 1998
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
Print No.: SPR2-230.033.01.01.02
Doc. Gen. Date: 07.98
General
Safety Information

CAUTION
Ensure compliance with the product-specific safety information in the documents as well as the general safety information found in ARTD, part 2.
Tests or settings performed using radiation are indicated by the radiation symbol ♠. Radiation protection clothing must be worn when performing such procedures.
Protective Conductor Measurement
After all work has been completed and all covers have been attached, perform a protective conductor measurement according to ARTD-002.731.17...
The protective conductor resistance must not exceed 0.2 Ohm.
Additional note
These instructions apply only to connecting a laser camera to a SIREMOBIL Iso-C.
Parts required
- Laser camera connection kit
37 79 159 G5429
Measurement devices and auxiliary materials required
• Service PC
• Digital Multimeter DMM
• Service Software (from logbook)
• Cable for connecting service PC to host 99 00 440 RE999
• Protective conductor meter
44 15 899 RV 090
Tools required
• Standard tool kit (refer to ARTD / section 3)
• 1 set of Allen keys
• Heat gun
• Drill
• 17 mm drill bit
• 16 mm drill bit
• 3 mm drill bit