Hamilton-T1 Technical Specifications

Product specifications 8 páginas · 396.76 KB · PDF

HAMILTON-T1

Technical specification for SW 2.2.x

Ventilation modes

Mode formMode nameModeAdult/PedNeonatal
Volume-targeted modes, adaptive pressure controlledAPVcmv / (S)CMV+Breaths are volume targeted and mandatory.✓✓
APVsimv / SIMV+Volume-targeted mandatory breaths can be alternated with pressure-supported spontaneous breaths.✓✓
Pressure-controlled modesPCV+All breaths, whether triggered by the patient or the ventilator, are pressure-controlled and mandatory.✓✓
PSIMV+Mandatory breaths are pressure controlled. Mandatory breaths can be alternated with pressure-supported spontaneous breaths.✓✓
DuoPAPMandatory breaths are pressure controlled. Spontaneous breaths can be triggered at both pressure levels.OO
APRVSpontaneous breaths can be continuously triggered. The pressure release between the levels contributes to ventilation.OO
SPONTEvery breath is spontaneous, with or without pressure-supported spontaneous breaths.✓✓
Intelligent ventilationASVOperator sets %MinVol, PEEP, and Oxygen. Frequency, tidal volume, pressure, and I:E ratio are based on physiological input from the patient.✓--
Noninvasive modesNIVEvery breath is spontaneous.OO
NIV-STEvery breath is spontaneous as long as the patient is breathing above the set rate. A backup rate can be set for mandatory breaths.OO
nCPAPDemand flow Nasal Continuous Positive Airway Pressure.--O
nCPAP-PCBreaths are pressure controlled and mandatory.--O
High flow oxygen therapyHiFlowO2High flow oxygen therapy. No supported breaths.OO

Standard: ✓ Option: O Not applicable: --

Standard configuration and options (in alphabetical order)

FunctionsAdult/PedNeonatal
Capnography, mainstream (volumetric) and sidestream○○
Communication ports: COM1 port, USB port, Nurse call○○
Communication protocols: for details see Connectivity brochure○○
Dynamic Lung✓--
Event log (up to 1000 events with data and time stamp)✓✓
Handles for various options (jet, ambulance, or bed rail mount)○○
IntelliTrig (leak compensation)✓✓
Languages✓✓
(English, US-English, Chinese, Croatian, Czech, Danish, Dutch, Finnish, French, German, Greek, Hungarian, Indonesian, Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Romanian, Russian, Serbian, Slovak, Spanish, Swedish, Turkish)✓✓
Manual breath/prolonged inspiration✓✓
NBC filter (only for HAMILTON-T1 Military)○○
Nebulization, pneumatic✓--
Night Vision Goggles (NVG) option○○
O2 enrichment✓✓
Oxygen adjustable from 21% to 100%✓✓
Patient group✓○
Print screen✓✓
Screen lock✓✓
Second battery (hot-swappable)○○
Speak valve compatibility○--
SpO2 monitoring○○
Standby with timer✓✓
Suctioning tool✓--
Trends/Loops○○
Flow trigger✓✓
Vent Status (Visual representation of ventilator dependence)✓✓

Standard: ✓ Option: o Not available: --

Technical performance

DescriptionSpecification
Automatic expiratory base flowAdult/Ped: Fixed at 3 l/min Neonatal: Fixed at 4 l/min
Inspiratory pressure0 to 60 cm $ H_{2}O $
Maximum inspiratory flow260 l/min (120 l/min with 100% O $ _{2} $)
Minimum expiratory time20% of cycle time; 0.2 to 0.8 seconds
Means of inspiratory triggeringFlow trigger control
Oxygen mixer accuracy$ \pm $ (volume fraction of 2.5% + 2.5% of actual reading)
Preoperational checksTightness test, Flow Sensor/O $ _{2} $ sensor/CO $ _{2} $ sensor calibration
Tidal volumeAdult/Ped: 20 to 2000 ml Neonatal: 2 to 300 ml
Brightness setting for displayThe range is 10% to 100% brightness. By default, Day is set to 80%; Night is set to 40%.
Brightness with NVG optionThe range is 1 to 10. The default is 5.
Standards and approvals
ClassificationClass IIb, continuously operating according to EC directive 93/42/EEC
CertificationIEC 60601-1:2005/A1:2012, IEC 60601-1-2:2014, ANSI/AAMI ES60601-1:2005/(R)2012, ISO 80601-2-12:2011, CAN/CSA-C22.2 No. 60601-1:14, EN ISO 5356-1:2015, ISO 80601-2-55:2011
DeclarationThe HAMILTON-T1 was developed in accordance with pertinent international standards and FDA guidelines. The ventilator is manufactured within an EN ISO 13485 and EN ISO 9001, Council Directive 93/42/EEC, Annex II, Article 3 certified quality management system. The ventilator meets the Essential Requirements of Council Directive 93/42/EEC, Annex I.
Electromagnetic compatibilityAccording IEC 60601-2:2014
Air transport compatibilityEN1789:2007+A2:2014, EN 13718-1:2014, EN 794-1:1997 +A2:2009, RTCA/DO-160G, MIL-STD-461F, MIL-STD-810G, Chg. 1, Meth 500.6, Proc. II Operation/Air Carriage, mod.
Safety ClassClass II, Type B applied part (ventilator breathing system, VBS), type BF applied part CO2 sensor including CO2 module connector; SpO2 sensor including adapter, continuous operation according to IEC 60601-1
Pneumatic performance
High-pressure O2Pressure:2.8 to 6 bar / 41 to 87 psi
Connector:DISS (CGA 1240) or NIST
Low-pressure O2Pressure:Maximum 6 bar / 87 psi
Connector:Quick-coupling system, compatible with Colder Products Company (CPC) PMC series
Air supplyIntegrated turbine
Inspiratory outlet (To patient port)Connector:ISO ID15/OD22 conical
Expiratory outlet (From patient port)Connector (on expiratory valve)ISO ID15/OD22 conical

Electrical specifications

Graphical patient data
Graphic type/tab nameOptions
WaveformsPressure, Volume, Flow, $ PCO2^{{1}} $, $ FCO2^{{1}} $, Plethysmogram $ ^{{2}} $
Intelligent panelsDynamic Lung $ ^{{3}} $, Vent Status, ASV Graph $ ^{{4}} $
Trends1-, 6-, 12-, 24-, or 72- $ h^{{5}} $ trend data for a selected parameter or combination of parameters
LoopsPressure/Volume, Pressure/Flow, Volume/Flow, Volume/PCO2 $ ^{{1}} $, Volume/FCO2 $ ^{{1}} $