R Series $ ^{®} $

Technical Specifications:

Image

21 117 143

Simple, Smart, and Code-Ready $ ^{®} $

The ZOLL® R Series Monitor/Defibrillator provides clinicians with comprehensive support for resuscitation. This includes cutting-edge technology to help you meet current guidelines for achieving high-quality CPR, as well as OneStep™ Resuscitation Electrodes that simplify and speed therapy. And to help ensure that the R Series will be Code-Ready, it conducts an automated self-test daily.

Driving High Quality CPR

• CPR Dashboard™ featuring Real CPR Help®—Guides rescuers with real-time audio and visual feedback on CPR quality measures. Provides numeric displays of depth and rate and visual indicators of compression release, as well as a unique perfusion performance index.

• See-Thru CPR®—Reduces the duration of pauses during CPR by filtering out the CPR artifact so rescuers can see whether an organized underlying rhythm has developed without stopping compressions.

Comprehensive Readiness Checks

The R Series® extends testing beyond the basic test shock to checking more than 100 individual indicators of readiness, including electronics, batteries, cables, electrodes, and defibrillator discharge. A green check mark or a red X on the defibrillator indicates the status of the device, as indicated during its last automated self-test. If a problem is detected, the R Series will turn on the display and an alert will be visible. When used with OneStep electrodes, even the electrodes are tested automatically every 24 hours.

Unmatched Clinical Excellence

• Unique, constant-current 40-msec pacing has the highest capture rate at the lowest average current required, assuring efficacy and patient comfort. $ ^{1} $

• Our high-current Rectilinear Biphasic™ waveform delivers the highest constant current at the optimal duration for defibrillation.

General

Size: 8.2 in (20.8 cm) high x 10.5 in (26.7 cm) wide x 12.5 in (31.7 cm) deep.

Weight: 13.6 lbs (6.2 kg) with OneStep Cable and SurePower™ Battery pack; 15.2 lbs (6.9 kg) with paddles.

Power Sources: AC Mains: 100 to 120 V AC (50/60 Hz), 220 to 240 V AC (50 Hz); Battery: Rechargeable lithium ion battery pack.

Low Battery Indicator: A "LOW BATTERY" message appears on the monitor when there is less than 15 minutes of ECG monitoring.

Design Standards: Meets or exceeds applicable requirements of UL 60601, AAMI DF80, IEC 60601-2-4, EN 60601-2-25, and 60601-2-27.

Patient Safety: All patient connections are electrically isolated.

Environmental: Operating Temperature: 0°C to 40°C, Storage and Shipping Temperature: -20°C to 60°C, Humidity: 5% to 95% relative humidity, noncondensing; Vibration: IEC 68-2-6 and IEC 68-2-34; Shock: IEC 68-2-27, 50 g/m³ half-sine; Operating Pressure: 594 to 1060 millibars; Particle and Water Ingress: IEC 529, IP22; Electromagnetic Compatibility (EMC): CISPR 11 Class B Radiated and Conducted Emissions; Electromagnetic Immunity: AAMI DF80, EN 61000-4-3 to 10 V/m; Electrostatic Discharge: AAMI DF80, EN 61000-4-2; Conducted Susceptibility: EN 61000-4-4, 61000-4-5, 61000-4-6.

Image

High-Current Rectilinear Biphasic Waveform

Image

Depth Rate CPR 2.4 In cpm Release PPI

CPR Dashboard

Featuring Real CPR Help

ADVANCING RESUSCITATION. TODAY.

ZOLL Medical Corporation

Worldwide Headquarters

269 Mill Road

Chelmsford, MA 01824

978-421-9655

800-348-9011

For subsidiary addresses and fax numbers, as well as other global locations, please go to www.zoll.com/contacts.

Defibrillator

Patient Impedance Range: 15 to 300 ohms. Energy Selections: 1 to 10, 15, 20, 30, 50, 75, 100, 120, 150 and 200 joules selected using controls on front of the defibrillator or sternum paddles. (Note: When using appropriate pediatric resuscitation electrodes, the 75-joule setting is replaced by 70- and 85-joule settings.)

Smart Step Energy Levels: Automatically escalates energy through a configured adult or pediatric protocol.

Energy Display: Shown on monitor for both selected and delivered.

Charge Time: Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules); longer charge times may result with a depleted or older battery.

Synchronized Mode: Synchronizes defibrillator pulse to patient's R wave. "SYNC" message displayed on monitor and markers shown on both monitor and recorded ECG.

Charge Controls: Control from front of defibrillator or apex paddle.

Paddles: External apex/sternum paddles; adult plates slide off to expose pediatric electrode surface.

Code Readiness Testing: Verifies defibrillator hardware, therapy delivery