Saltar al contenido principal

ResMed AirSense 10 Top Cover Replacement

Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.

Qué necesitas

  1. ResMed AirSense 10 Top Cover Replacement, Top Cover: paso 1, imagen 1 de 2 ResMed AirSense 10 Top Cover Replacement, Top Cover: paso 1, imagen 2 de 2
    • Remove the water tub by gently pulling it out.

  2. ResMed AirSense 10 Top Cover Replacement: paso 2, imagen 1 de 3 ResMed AirSense 10 Top Cover Replacement: paso 2, imagen 2 de 3 ResMed AirSense 10 Top Cover Replacement: paso 2, imagen 3 de 3
    • Pry off the face plate by using your finger or an opening tool.

    • Use a moderate amount of force when prying open.

    • Remove the face plate from the top cover.

    • There will be a clicking sound on either side of the divot when assembling and disassembling.

  3. ResMed AirSense 10 Top Cover Replacement: paso 3, imagen 1 de 2 ResMed AirSense 10 Top Cover Replacement: paso 3, imagen 2 de 2
    • Unscrew and remove the two 11.8 mm TR10 Torx screws that were revealed from the face plate being removed.

  4. ResMed AirSense 10 Top Cover Replacement: paso 4, imagen 1 de 2 ResMed AirSense 10 Top Cover Replacement: paso 4, imagen 2 de 2
    • Unscrew and remove the two 11.8 mm TR10 Torx screws from the bottom of the device.

  5. ResMed AirSense 10 Top Cover Replacement: paso 5, imagen 1 de 3 ResMed AirSense 10 Top Cover Replacement: paso 5, imagen 2 de 3 ResMed AirSense 10 Top Cover Replacement: paso 5, imagen 3 de 3
    • Remove the top cover from the rest of the device.

Conclusión

To reassemble your device, follow these instructions in reverse order.