Puritan Bennett 560 Exhalation Block Replacement

Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.
  1. Puritan Bennett 560 Exhalation Block Replacement, Exhalation Block: paso 1, imagen 1 de 1
    • Locate the exhalation block on the front of the ventilator.

  2. Puritan Bennett 560 Exhalation Block Replacement: paso 2, imagen 1 de 1
    • Tilt the machine up and unscrew the knob that holds the exhalation block in place.

    • This knob can be untwisted by hand, no tools required.

  3. Puritan Bennett 560 Exhalation Block Replacement: paso 3, imagen 1 de 1
    • Pinch the "From Patient" fitting and slide the exhalation block away from the ventilator.

  4. Puritan Bennett 560 Exhalation Block Replacement: paso 4, imagen 1 de 1
    • After replacing or cleaning the block, perform the exhalation flow sensor calibration.

    • See 6.7.3.3 in the Service Manual for instructions on calibration.

Conclusión

To reassemble your device, follow these instructions in reverse order.