    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. How to Disconnect a Philips Respironics V60: paso 4, imagen 1 de 1
    • Disconnect the internal battery from the power harness connector.

Conclusión

To reassemble your device, follow these instructions in reverse order.