How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff

Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.
    • Adjust the gas flow to 10 L/min.

  1. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: paso 2, imagen 1 de 1
    • Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.

  2. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: paso 3, imagen 1 de 1
    • Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20

  3. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: paso 4, imagen 1 de 1
    • Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.

Conclusión

Alternate settings for the Max Pressure Relief should be made in accordance with hospital protocol.