Qué necesitas

  1. Hamilton-T1 Oxygen Cells Replacement, Oxygen Cells: paso 1, imagen 1 de 1
    • Flip device around until the back panel is exposed.

  2. Hamilton-T1 Oxygen Cells Replacement: paso 2, imagen 1 de 1
    • Remove the screw in the bottom center of the panel, and then take the panel off.

  3. Hamilton-T1 Oxygen Cells Replacement: paso 3, imagen 1 de 1
    • Disconnect the oxygen cell by removing the tube.

    • Then twist the cell counter-clockwise and pull to remove the oxygen cell.

    • Install the new oxygen cell and put the back cover on.

    • After the cell is installed. Run the oxygen cell calibration.

    • Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.

Conclusión

After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.