Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement

Qué necesitas

  1. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement, Opening the Column in the Hamilton Galileo: paso 1, imagen 1 de 1
    • Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.

    • Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).

  2. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement: paso 3, imagen 1 de 1
    • Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.

  3. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement: paso 4, imagen 1 de 2 Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement: paso 4, imagen 2 de 2
    • Without removing any electrical connections, open the left side of the column as wide as practical.

    • This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.

    • Make sure no electrical connections are touching the column.

    • Make sure all green/brown grounding (earthing) cables are securely in place.

  4. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement, Control Board and GMP Assembly: paso 6, imagen 1 de 1
    • Locate the control board and GMP assembly.

    • Remove all cable connectors joining the control board and the GMP board to other parts of the GALILEO.

  5. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement: paso 8, imagen 1 de 1
    • Remove the single grounding (earthing) screw securing the control board to the aluminium mounting plate.

  6. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement: paso 9, imagen 1 de 1
    • Remove the control board from the plastic spacers. Do do this, squeeze each spacer in turn, while easing the board off the spacer. Leave the spacer in the center of the board until last.

  7. Hamilton Galileo Ventilator Control Board and GMP Assembly Replacement: paso 10, imagen 1 de 1
    • To re-mount the control board, reverse steps seen here. Only press the control board onto plastic spacers, do not use pliers. See image for correct cable layout.

Conclusión

To reassemble, mount the control board back onto plastic spacers without using pliers and be sure to connect all cables together correctly.

Jacob Dixon

Miembro Desde 04/09/20

191 Reputación

1 Guía escrita

Equipo

Cal Poly, Team S7-G26, Paton Spring 2020 Miembro de Cal Poly, Team S7-G26, Paton Spring 2020

CPSU-PATON-S20S7G26

3 Miembros

4 Guías creadas

0 comentarios

Agregar Comentario

Ver Estadísticas:

Ultimas 24 horas: 0

Ultimos 7 días: 0

Ultimos 30 días: 6

Todo El Tiempo: 335