GE Carescape R860 Alarm Light (PWA) Board Replacement

Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.

Qué necesitas

  1. GE Carescape R860 Alarm Light (PWA) Board Replacement, GE Carescape R860 Display Unit Disassembly: paso 1, imagen 1 de 1
    • Using an opening tool, gently pry the alarm light lens (boxed) from the display unit.

  2. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 2, imagen 1 de 1
    • Loosen the seven screws that hold the rear housing using a Phillips #0 screwdriver.

    • Remove the rear housing.

  3. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 3, imagen 1 de 1
    • Cut the cable tie holding the two cables at the lower right side using scissors.

  4. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 4, imagen 1 de 1
    • Disconnect the touch controller cable.

  5. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 5, imagen 1 de 1
    • Disconnect the Keypad harness at the Carrier Board.

  6. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 6, imagen 1 de 1
    • Disconnect the encoder harness from the Carrier Board.

  7. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 7, imagen 1 de 1
    • Disconnect the speaker harness from the Carrier Board and set the speakers aside.

  8. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 8, imagen 1 de 1
    • Disconnect the alarm light harness from the Carrier Board.

  9. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 9, imagen 1 de 1
    • Loosen (do not remove) the five screws at the top edge of the shield just enough to raise it and slide the shield away with a Phillips #0 screwdriver.

  10. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 10, imagen 1 de 1
    • Loosen the four screws along each side of the shield with a Phillips #0 screwdriver.

  11. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 11, imagen 1 de 1
    • Remove the four screws at the bottom edge of the shield with a Phillips #0 screwdriver.

  12. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 12, imagen 1 de 1
    • Slide the shield down to disengage the touch controller harness connector.

  13. GE Carescape R860 Alarm Light (PWA) Board Replacement, Alarm Light (PWA) Board: paso 13, imagen 1 de 1
    • Pull the PWA board from the slot.

    • Replace the PWA board.

  14. GE Carescape R860 Alarm Light (PWA) Board Replacement: paso 14, imagen 1 de 1
    • Reconnect the alarm light harness.

Conclusión

To reassemble your device, follow these instructions in reverse order.

Andre Fisher

Miembro Desde 01/23/20

357 Reputación

2 Guías creadas

Equipo

Cal Poly, Team S7-G9, Paton Spring 2020 Miembro de Cal Poly, Team S7-G9, Paton Spring 2020

CPSU-PATON-S20S7G9

3 Miembros

4 Guías creadas

0 comentarios

Agregar Comentario

Ver Estadísticas:

Ultimas 24 horas: 0

Ultimos 7 días: 0

Ultimos 30 días: 1

Todo El Tiempo: 265