Saltar al contenido principal

Access to Housing in the Ulco Integrus PSV

Equipo médico
El fabricante de este equipo médico no ha avalado o contratado a iFixit ni a esta guía de reparación.
    • Before disassembly, make sure the front wheel breaks are on.

    • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

  1. Access to Housing in the Ulco Integrus PSV: paso 4, imagen 1 de 1
    • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

  2. Access to Housing in the Ulco Integrus PSV: paso 5, imagen 1 de 1
    • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

Conclusión

To reassemble your device, follow these instructions in reverse order.